Last update May 11 2009. |
|
| Calibration temperature range | Least uncertainty of calibrations (Best measurement) |
For loggers without probes fitted (Internal sensor) | |
| -15ºC to 40°C | ± 0.061°C (routinely ±0.2°C from -15 to 40°C) in the F3 machine. |
For loggers with probes fitted (External sensor) |
|
| -24°C to 130°C | ±0.056°C (routinely ±0.2°C from -20 to 50°C) in dry block calibrator. ±0.036°C (routinely ±0.2°C from -20 to 50°C) in stirred liquid baths. |
Single point temperature calibrations (new) and re-calibrations to loggers |
|
| Calibration temperature range | Least uncertainty of calibrations ( Best measurement ) |
For loggers without probes fitted (Internal sensor) | |
| -24°C to 80°C | ±0.061°C in stirred liquid baths with comm block fitted. |
For loggers with probes fitted (External sensor) |
|
| -24°C to 130°C | ±0.056°C At single temperature points in dry block calibrator. ±0.036°C At single temperature points in stirred liquid baths |
Relative Humidity calibrations (new) and re-calibrations to loggers |
||
| Calibration humidity range | Least uncertainty of calibrations (Best measurement) | |
Single humidity points between 12% and 85% RH |
||
| 12% to 85% RH | ±1.0% RH at a dry bulb temperature between 20°C and 30°C and a dew point range of -6°C to 23°C | |
Calibration over the entire range from 12% to 85% RH |
||
| 3.0% RH at a dry bulb temperature between 20°C and 30°C and a dew point range of -6°C to 23°C | ||