Safe range limit (LCD logger only)

Contents  Previous  Next

Program Menu > Safe range limit (LCD logger only)

The safe range limit sets the number of samples that must elapse before the logger has entered the safe range before a safe range entry error condition is signalled.  In other words, if the safe range limit is set to 20 samples, then the logger must have reached the safe range before 20 samples have been taken.

 

This parameter is only available in the G4 LCD logger.  The Mk3 Temprecord loggers also implement a safe range delay, but the value of the delay (in samples) is shared with the limit delay, i.e. setting the limit delay in the Mk3 logger also sets the safe range entry delay, and the limit delay in turn is shared between the temperature and humidity channels, whereas in the G4 (LCD) logger these parameters are separate and can also be independently programmed for each humidity or temperature channel.

 

 