The temperature exceeded the limits but the summary does not show this

Contents  Previous  Next

Common Problems > The temperature exceeded the limits but the summary does not show this

You may have the limit delay parameter on the logger set to a non-zero value.  If the limit delay is set to 5 for example, the logger will allow 5 consecutive samples outside the limits before it reports an out-of-limits condition.

 

If you want the logger to report an out-of-limits condition as soon as a sample is outside the limits, set the limit delay to zero.