Auto Mode Options

Contents  Previous  Next

Options Menu > Auto Mode Options

Use the Options/Auto Mode page to change the behaviour of Temprecord’s Auto Mode.  Auto Mode providfes a means of rapidly processing a quantity of loggers, with little more user intervention than inserting and removing the logger from a reader.

 

Timeout

This option sets the number of seconds before auto mode times out and exits when either waiting for a logger or waiting for the logger to be removed after completion.

 

Stop Logger if necessary

Check this option if you want Temprecord Auto Mode to stop a logger that is logging.  If the logger is already stopped, no action is taken

 

Read Logger

Check this option if you want Temprecord Auto Mode to read the data from the logger.

 

Save TRX file

Check this option if you want Temprecord Auto Mode to read the data from the logger and save it to a disk file.  The name used for the file is determined by the formatted filename option.  The folder used is determined by the folder to save TRX files.  If you select the Save TRX file option, the Read Logger option will be selected for you.

 

Save to Web (TRX files)

Check this option if you want Temprecord Auto Mode to read the data from the logger and save it to Temprecord's remote web storage facility.  You must have arranged this facility beforehand with Temprecord.  The name used for the file is determined by the formatted web filename option.  The folder used is determined by the folder to save web TRX files.  If you select the Save to Web option, the Read Logger option will be selected for you.

 

Email TRX file

Check this option if you want Temprecord Auto Mode to email the TRX file.  If you select the Email TRX file option, the Read Logger option will be selected for you, and the data is automatically saved to a TRX file, regardless of the Save TRX file auto mode option.

 

Save PDF File

Check this option if you want Temprecord Auto Mode to read the data from the logger and save it to a PDF file.  The name used for the file is determined by the formatted PDF filename option.  The folder used is determined by the folder to save PDF files.  If you select the Save PDF File option, the Read Logger option will be selected for you.

 

The contents of the PDF report (summary, graph, etc) are determined by the printing options.  The printed pages range option is ignored when generating the PDF.  All pages of the report are generated to the PDF, even if a page range was selected in the print to PDF dialog.

 

Save to Web (PDF files)

Check this option if you want Temprecord Auto Mode to read the data from the logger, genmerate a PDF report file, and save it to Temprecord's remote web storage facility.  You must have arranged this facility beforehand with Temprecord.  The name used for the file is determined by the formatted web PDF filename option.  The folder used is determined by the folder to save web PDF files.  If you select the Save to Web option, the Read Logger option will be selected for you.

 

Email PDF File

Check this option if you want Temprecord Auto Mode to email the PDF report.  If you select the Email PDF File option, the Read Logger option will be selected for you, and the data is automatically saved to a PDF file, regardless of the Save PDF File auto mode option.

 

Print Report

Check this option if you want Temprecord Auto Mode to send the printed report to the printer.  The contents of the printed report (summary, graph, etc) are determined by the printing options.  All pages of the report are printed, even if a page range was selected in the print dialog. If you select the Print Report option, the Read Logger option will be selected for you.

 

Reuse Logger

Check this option if you want Temprecord to reuse the logger. The logger must be stopped. Remember that reusing a logger clears any logged data so you should use the Read Logger and Save TRX file options in conjunction with Reuse Logger also.

 

Reusing the logger will only be successful if the logger is in the Stopped state by the time the Reuse Logger stage of Auto Mode is reached.

 

Load Parameters

Check this option if you want Temprecord Auto Mode to program a set of parameters into the logger. The parameter set programmed can be those parameters as set in the Default Parameter settings, or they can come from a Named Parameter Set.

 

Note that if this function is used, whatever parameters are programmed into the logger will be overwritten, including the user data, which may have been set explicitly for the previous trip.

Loading parameters will only be successful if the logger is in the Ready state by the time the Load Parameters stage of Auto Mode is reached (for example, if the logger was stopped and then read by previous Auto Mode steps, then the Reuse Auto Mode step will need to be enabled).

 

Start Logger

Check this option if you want Temprecord Auto Mode to start the logger.  

 

Starting the logger will only be successful if the logger is in the Ready state by the time the Start Logger stage of Auto Mode is reached (for example, if the logger was stopped and then read by previous Auto Mode steps, then the Reuse Auto Mode step will need to be enabled).

See also:

Auto Mode Operation

Named Parameter Sets

Default Options

Formatted File and Folder Names

Formatted Web File and Folder Names

 

 


(C) 1992-2019 Temprecord International Ltd
12/10/2017 12:46:18 PM
Temprecord Web Help